Barovue™10 is a highly accurate barometer that can measure pressures over a range of 500 to 1100 hPa and can be used in a range of applications that includes meteorology and hydrology. This sensor includes our zero downtime recalibration. Other sensors on the market need to be taken offline and sent back to the manufacturer to be recalibrated, often for weeks at a time, causing a loss in valuable data and time, as well as an additional expense.
The sensor card for the BaroVUE™10 is pre-calibrated. Replacement sensor cards are also pre-calibrated, requiring only one trip to the field station and no downtime. The design of the BaroVUE™10 allows you to remove and replace your sensor card quickly and easily without either having to disassemble the equipment in the enclosure or having to send the equipment back to the manufacturer.
Read MoreBarovue™10 is a low-power, digital barometer (-40 to +60°C) that provides best-in-class accuracy and stability (±0.1 hPa/yr) over the entire pressure and temperature range. It can run continuously and does not require power cycling unlike other products in this category. The digital output is both SDI-12 and RS-232, which reduces noise and measurement uncertainty compared with analog sensors, and makes the BaroVUE™10 compatible with all Campbell Scientific SDI-12 devices. Moreover, this barometric pressure sensor can be installed and integrated into many new and existing weather stations. The transducers used in the BaroVUE™10 are direct-to-digital sensors, and no reconversion takes place inside the barometer.
-笔记- | These specifications assume the recommended desiccation is used. |
压力范围 | 500 to 1100 hPa |
温度范围 | -40°至 +60°C |
电源电压 | 9 to 28 Vdc |
海拔 | 海拔〜609.6 m(2,000 ft)以下海平面至4,572 m(15,000英尺) |
目前的消费 |
|
数字输出 | SDI-12,RS-232系列 |
压力拟合 | 带刺的安装为0.318 cm(0.125英寸) |
Dimensions | 2.2 x 9.0 x 10.2 cm (0.87 x 3.54 x 4.02 in.) |
Weight | 226.8 g(0.5磅) |
准确性 |
|
Calibration Uncertainty | ±0.15 hPa |
不确定 | ±0.3 hPa(在20°C下) |
Total Uncertainty | ±0.5 hPa (at -40° to +60°C) |
Long-Term Stability | ±0.1 hpa/yr |
电源拒绝 | 微不足道 |
测量噪声 | 0.05 HPA(RMS) |
解析度 | 0.1 hPa |
Note:以下显示了显着的兼容性信息。这不是所有兼容或不兼容产品的全面列表。
Product | Compatible | Note |
---|---|---|
ENC100 | ||
ENC10/12 | ||
ENC10/12R | ||
ENC12/14 | ||
ENG14/16 | ||
ENG16/18 | ||
ENC16/18UL | ||
ENG24/30 | ||
ENG24/30S | ||
ENG24/30SUL | ||
ENC24/30UL | ||
ENG8/10 |
Product | Compatible | Note |
---|---|---|
DAQ案例15 |
描述:包括一个使用的CR1000X程序C1!SDI-12命令检索数据。该计划将使气压测量值调整为海平面。调整在程序开始时以常数的高程_M输入。最初,高程_M值为1382;以米为单位将此值更改为您的站点高程。
Number of FAQs related toBaroVUE10:13
展开全部全部收缩
当使用SDI-12或RS-232通信协议时,可以在HPA,KPA,MMHG,INHG或PSI中返回压力测量。参考Barovue™10手册以获取更多信息。如果需要不同的测量单元,请参阅转换图表。
可能正在使用旧版本的捷径。下载最新版本的捷径。
联系您当地的坎贝尔科学办公室购买校准的Barovue™10传感器卡(BVC10)。收到校准卡后,用当前卡将其交换,然后将其邮寄回您当地的坎贝尔科学办公室。这样,只需要一次前往现场站点的旅行,传感器将不存在停机时间。
Campbell Scientific建议每年进行校准。但是,符合已发布的标准或法规可能需要更频繁地。
虽然测量值等效,但Barovue™10不是替换。与CS100和CS106不同,Barovue™10是通过SDI-12或RS-232通信的数字传感器。使用Barovue™10 in a system that previously used a CS100/CS106, the wiring and program will need to be changed, as it is a digital sensor.
零停机时间/重新校准
Barovue™10旨在在正常使用过程中允许零停机时间重新校准。市场上的其他传感器需要离线并将其送回制造商进行重新校准,通常每次一次数周,从而导致数据损失以及时间和金钱。
In comparison, the sensor card of the BaroVUE™10 is calibrated ahead of time and sent out to customers to replace the old sensor card, requiring only one trip to the field and no downtime.
除此之外,Barovue的设计™10允许客户快速轻松地卸下和更换其传感器卡,而无需拆卸外壳中的设备或必须将产品发送回制造商。无需关闭设备即可更换卡。
高度可靠的测量
Barovue™10传感器卡使用统计分析的多个传感器单元,以产生准确,精确且可重复的测量以及数据质量度量。客户可以使用此质量指标来跟踪校准周期,规划,维护时间表和检测到的过失条件。
低电量
Barovue™10是低功率设备,可以保持连续运行。Barovue™与此类别中的其他产品不同,10不需要任何电源循环。
非常低的噪音
Barovue™10使用直接到数字的传感器,并且气压计内部没有进行类似于数字的重新转换。
These factors, coupled with the multiple sensor cells, result in a reduction in noise when compared with analog sensors.
坎贝尔科学美学和形式
Barovue™10 has been designed with functionality in mind, and its aesthetic complements other Campbell Scientific products.
Barovue™10 can be mounted in the orientation that best suits the application and has a smaller backplane footprint than its competitors.
Barovue™10大,可以容纳一个ENC100。为铁道部e information, refer to the Specifications information on theENC100产品页面。
Barovue没有热身时间™10。
质量度量是一个独特的诊断功能,表明传感器卡返回的压力测量的质量(BVC10)。它使用户能够监视其传感器的健康状况,以更好地安排重新校准。
质量度量范围 | 描述 | 评论 |
> 6 |
传感器新鲜校准 |
|
5至5.99 |
Sensor working well |
|
4 to 4.99 |
In two to three months, this sensor will need to be recalibrated. |
在三个月内,发送Barovue™10到工厂进行重新校准。 |
0至3.99 |
现在重新校准传感器。 |
尽快发送Barovue™10到工厂进行重新校准。 |
In most cases, the enclosures or buildings where the BaroVUE™10 is housed have enough ventilation for the pressure to stay at equilibrium with the ambient pressure. Therefore, the sensor does not need to be vented to the outside
However, if the sensor resides in an air-tight environment, and the pressure external to this environment needs to be measured, then the sensor will need to be vented to the outside. The preferred method for measuring the pressure external to the enclosure is to use a two-plate pressure port such as the RM Young 61002 to minimize dynamic pressure errors due to wind.
We've updated our privacy policy.Learn More